A biomechanical comparison of the two- and four-hole side-plate dynamic hip screw in an osteoporotic composite femur model

22

Citations

24

References

2017

Year

Abstract

The results of this study demonstrate that DHS constructs with a two- or four-hole side-plate are biomechanically comparable with regard to axial and torsional stiffness and load to failure in an osteoporotic composite femur model. In a healed intertrochanteric fracture model, a two-hole construct did not appear to be more prone to peri-implant fracture. To date, a biomechanical comparison of these two implants with regard to torsional forces has not been reported.