The Incidence of Dissociated Liners in 4,751 Consecutive Total Hip Arthroplasties Using Pinnacle Polyethylene Acetabular Liners

22

Citations

24

References

2017

Year

Abstract

Ensuring correct liner seating/locking, minimising impingement, achieving appropriate component version and avoiding radiographic cup inclinations >50° should minimise the risk of liner dissociation. Any new noise or squeaking from a polyethylene liner should undergo radiographic investigation to exclude dissociation. We recommend managing late cases of liner dissociation with revision of the acetabular shell if the cup orientation could be improved or if there is any damage to the liner-locking groove, to reduce the risk of recurrence.