Publication | Open Access
Inappropriate shock in a subcutaneous cardiac defibrillator due to residual air
11
Citations
9
References
2017
Year
Cardiogenic ShockCardiopulmonary ResuscitationCardiac AnaesthesiaSubcutaneous Cardiac DefibrillatorBlunt Finger DissectionResidual AirPatient SafetyCardiovascular ImagingInappropriate ShockSurgeryElectrophysiologyTraumatic Cardiac ArrestMedicineCardiologyEmergency MedicineCardiac Arrest
Inappropriate shock due to residual air in subcutaneous implantable cardiac defibrillators is not a well-known complication. Obtaining overpenetrated X-rays, recognizing electrocardiogram findings, limiting blunt finger dissection, and switching to sense at another vector are techniques which might lead to avoidance of unnecessary wound exploration or device removal.