Inter-instrumental comparison for the measurement of electrolytes in patients admitted to the intensive care unit

11

Citations

7

References

2017

Year

Abstract

It can be concluded that AA and ABG analyzers may be used interchangeably for measurement of potassium in the Institute of Medicine, while the same cannot be concluded for the measurement of sodium, because of the significant difference in sodium measurement by the two instruments.