MRD detection in multiple myeloma: comparison between MSKCC 10-color single-tube and EuroFlow 8-color 2-tube methods

Mikhail Roshal, Juan Flores‐Montero, Qi Gao, Maesa Koeber, Jessica Wardrope, Brian G.M. Durie, Ahmet Doǧan, Alberto Órfão, Ola Landgren

Blood Advances · 2017 · 116 citations · 8 references

DOIFull text

Open access

Concepts

Abstract

= 0.97). The overall concordance between the 2 tests was 98%. In 1 case, there was a discrepancy near the limit of detection of both tests in favor of the slightly greater theoretical sensitivity of the EuroFlow 8-color 2-tube panel (analytical sensitivity limit of MSKCC single 10-color tube: 6 cells in 1 million with at least 3 million cell acquisitions; EuroFlow 8-color 2-tube panel: 2 cells in 1 million with the recommended 10 million cell acquisitions).