Journal of Instrumentation · 2017 · 17 citations · 7 references
MedicineRadiation OncologyRadiation DosimetryRadiation MeasurementTreatment PlanningDose CalibrationDosimetry InstrumentationPatient-specific DosimetryQuality Assurance In DosimetryRadiation Dose MeasurementRadiation TherapyIntensity-modulated Radiation Therapy DosimetryIntensity-modulated Radiation Therapy Quality AssuranceMeasurement ReproducibilityPercentage Depth DoseRadiotherapy Quality AssuranceTherapeutic DosimetryTreatment Planning DosimetryX-ray Dosimetry
In radiation therapy, accurate radiotherapy treatment plan (RTP) reproduction is necessary to optimize the clinical results. Thus, attempts have recently been made to ensure high RTP reproducibility using image-guide radiation therapy (IGRT) technology. However, the clinical use of digital X-ray equipment requires extended quality assurance (QA) for those devices, since the IGRT device quality determines the precision of intensity-modulated radiation therapy. The study described in this paper was focused on developing a multi-energy PbO dosimeter for IGRT device QA. The Schottky-type polycrystalline PbO dosimeter with a Au/PbO/ITO structure was evaluated by comparing its response coincidence, dose linearity, measurement reproducibility, linear attenuation coefficient, and percent depth dose with those of Si diode and standard ionization chamber dosimeters.
7
Safety considerations for IGRT: Executive summary
David A. Jaffray, K Langen, G Mageras et al. · Practical Radiation Oncology · 2013 · 69 citations · Full text