Publication | Closed Access
<scp>FMEA</scp> of manual and automated methods for commissioning a radiotherapy treatment planning system
26
Citations
21
References
2017
Year
An FMEA of treatment planning commissioning tests using automation and standardization via API scripting, preloaded, and pre-modeled standard beam data, and digital phantoms suggests that errors and risks may be reduced through the use of an ACP.