We present the case of unanticipated airway ignition during hard palate biopsy. Transnasal humidified rapid-insufflation ventilatory exchange (THRIVE) and monopolar diathermy were utilised for the procedure, during which an arc arose from the diathermy tip to a titanium implant, causing a brief ignition on the monopolar diathermy grip. This case highlights the need for maintained awareness of fire risk when using diathermy in the presence of THRIVE during airway surgery.