DentistryImplant DentistryDental MaterialsMechanical Behavior Of MaterialsMechanical Properties Of Dental MaterialsImplant BiomechanicsImplant MaterialsAxial Load ResponseAxial Load TransferBiomechanical Testing Of ImplantsBiomechanics Of Dental ImplantsDental Implant BiomechanicsImplant Angulation EffectsStrain Gauge MeasurementWax Pattern Adaptation
Objective: This in vitro study was to evaluate micro-strains around of implant, under the influence of angulations and vertical misfit in three-element implant-supported fixed partial dentures during axial loading by using strain gauge analysis. Materials and Methods: Three external hexagon implants with straight configuration and three external hexagon implants with angled (17°) configuration were inserted into two polyurethane blocks. To measure micro-strain, four strain gauges were bonded onto the surface of each block. Plastic copings were adapted to a standard wax pattern and cast. An axial load of 30 kgf was applied on the center of each implant for 10 seconds, using a load application device. The vertical misfit was measured at six different points by using a stereo microscope with 100-X magnification. Results: The results showed that the values for different implant angulations were significant (P= 0.0086). The Pearson’s correlation test between micro-strain and vertical misfit revealed no correlation between angled configuration (P= 0.891) and straight configuration (P= 0.568). Conclusion: The micro strain was higher for angled implants; no correlation was found between the vertical misfit and the micro strain values.