Bioanalysis · 2017 · 54 citations · 20 references
PharmacyPharmaceutical AnalysisPharmacologyClinical BiochemistryClinical ImmunologyImmunopharmacologyBioanalytical Method ValidationBiomarker ValidationImmunosuppressionBiologic ImmunosuppressantsBiomarkers Of ImmunosuppressionClinical Data ValidationDried Blood Spot AnalysisDried Blood Spot SamplingHematocrit EffectsHome SamplingSampling
AIM: Hematocrit (Ht) effects remain a challenge in dried blood spot (DBS) sampling. The aim was to develop an immunosuppressant DBS assay on two LC-MS/MS systems covering a clinically relevant Ht range without Ht correction. RESULTS: The method was partially validated for tacrolimus, sirolimus, everolimus, cyclosporin A and fully validated for mycophenolic acid on an Agilent and Thermo LC-MS/MS system. Bias caused by Ht effects were within 15% for all immunosuppressants between Ht levels of 0.23 and 0.48 l/l. Clinical validation of DBS versus whole blood samples for tacrolimus and cyclosporin A showed no differences between the two matrices. CONCLUSION: A multiple immunosuppressant DBS method without Ht correction, has been validated, including a clinical validation for tacrolimus and cyclosporin A, making this procedure suitable for home sampling.
20
H. Passing, W. Bablok · Clinical Chemistry and Laboratory Medicine (CCLM) · 1983 · 1.5K citations · Full text
Remco A. Koster, Jan‐Willem C. Alffenaar, Ben Greijdanus et al. · Talanta · 2013 · 140 citations