Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test
20
Citations
9
References
2017
Year
Very good and highly reproducible image quality could be achieved on a 1.5T MR-sim with a flexible coil setting as revealed by the standard ACR MRI phantom test. The flexible RT-setting compromised in image signal-to-noise ratio (SNR) compared to the normal DR-setting, and resulted in reduced low-contrast object detectability.