Publication | Open Access
Sleep Apnea Screening with a Contact:Free Under:the:Mattress Sensor
18
Citations
6
References
2016
Year
Sleep apnea is a highly prevalent yet under-diagnosed condition. This study tested a novel algorithm for sleep apnea screening with a contact-free system based on a piezo-electric sensor (PE system -EarlySense Ltd). The study population included 96 subjects who were referred to a sleep study, and underwent a full overnight polysomnography (PSG) in a sleep lab. 16 participants were diagnosed with severe sleep apnea, 18 with moderate, 30 with mild and 32 with no sleep apnea. All subjects were simultaneously measured with the PE system. Respiration waveform was analyzed to extract time and frequency domain features and calculated an internal index for the number of apnea/hypopnea events. It also used an algorithm for sleep wake detection which is described elsewhere. Based on the internal apnea index and the duration of sleep, the system classified the subjects into two groups: one above and one below an Apnea-Hypopnea-Index (AHI) of 15. The classification was compared to a PSG classification of a blinded sleep expert.