Biomedical sensors use capacitance measurement for a variety of detection applications and capacitance measurement in the tens of femto-farad range need a low-noise environment. The capacitance sensor interface converts the capacitance change to frequency change to mitigate the noise and to operate in a noisy environment. Using differential sensing architecture and calibration, the frequency domain technique allows measuring small capacitance values even in the presence of large parasitic capacitance. The impact of the voltage variation due to temperature or supply is removed by one-point calibration The test circuit was designed in 130-nm CMOS technology and the capacitance was measured for 1-100 fF having a nominal (or parasitic) capacitance of 1 pF.