Concepedia

Publication | Open Access

Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator

365

Citations

30

References

2017

Year

Abstract

In this study, device or lead failure did not occur in any patient with a non-MRI-conditional pacemaker or ICD who underwent clinically indicated nonthoracic MRI at 1.5 tesla, was appropriately screened, and had the device reprogrammed in accordance with the prespecified protocol. (Funded by St. Jude Medical and others; MagnaSafe ClinicalTrials.gov number, NCT00907361 .).