Radiation Protection Dosimetry · 1987 · 26 citations · 0 references
EngineeringMaterials ScienceNuclear EngineeringRadiation DetectionRadiation DosimetryRadiation MeasurementAbsorbed DoseDosimetry InstrumentationNeutron DetectionPersonal DosimetryRadiation Dose MeasurementRadiation Monitoring SystemsAlbedoAngular ResponseFast Neutron DetectionNeutron Detector SystemsNeutron DetectorsNeutron Personal DosimetryNeutron Radiation AssessmentThermal Neutron Detection
A simple system which can be used to assess rapidly the area density of chemically etched pits is described. Studies of its application to neutron dosimetry using CR-39 have recently been extended. It has been shown that the previously reported fast neutron low energy threshold of 180 keV can be reduced to below 140 keV. Various boron-coated foils have been used to demonstrate that the system can also be used to measure thermal neutron exposure over a range exceeding seven decades. It would therefore form the ideal detection system for an albedo dosemeter. Two or three fast neutron sensitive CR-39 elements set back at an angle of about 35o together with a thermal neutron sensitive albedo element would constitute a personal neutron dosimetry system with energy and angular response close to the ideal.