Concepedia

Publication | Open Access

Randomized, masked, in vitro comparison of three commercially available tear film osmometers

32

Citations

15

References

2017

Year

Abstract

In this randomized, masked, in vitro study, device A and device B had significantly better accuracy and precision in measuring osmolarity of contrived tear solutions of known target values compared to device C. Device C showed insufficient performance to accurately and precisely delineate osmolarity levels in the physiological range. Furthermore, in vivo studies would be required to compare performance in human subjects.