Concepedia

Publication | Open Access

Parathormone stability in hemodialyzed patients and healthy subjects: comparison on non-centrifuged EDTA and serum samples with second- and third-generation assays

17

Citations

15

References

2017

Year

Abstract

PTH seems more stable in EDTA than in serum gel tubes but only when samples have to stay unprocessed for a long period (18 h) at room temperature (25°C), which can happen when samples are delivered from external care centers. For all the other conditions, using serum gel tubes is recommended since calcium measurement, which is necessary for a good PTH results interpretation, can be achieved on the same tube.

References

YearCitations

Page 1