Concepedia

Publication | Open Access

Analysing a mechanism of failure in retrieved magnetically controlled spinal rods

35

Citations

18

References

2017

Year

Abstract

We found that a third of the retrieved magnetically controlled growth rods had failed due to pin fracture secondary to corrosion of the internal mechanism. We recommend that surgeons consider that any inability of magnetically controlled growth rods to distract may be due to corrosive debris building up inside the mechanism, thereby preventing normal function.

References