MedicineRadiation OncologyRadiologyRadiation PhysicsRadiobiologyTreatment PlanningExternal Beam RadiotherapyImage-guided RadiotherapyImage-guided TherapyParticle TherapyProton DosimetryProton TherapyImage-guided Particle TherapyImage-guided Proton TherapyProton Beam RadiobiologyProton Therapy DosimetryRadiobiological EffectivenessRadiobiological EndpointsRadiobiological Response
(10% peak dose). Alignment of the proton beam with the SARRP system isocenter was measured at 0.24 mm agreement. The width of the beam changes very little with depth. Monte Carlo-based calculations of dose using the CT image data set as input demonstrate in vivo use. Monte Carlo simulations of the modulated 100 MeV clinical proton beam show a significantly reduced Bragg peak. We demonstrate the feasibility of a proton beam integrated with a commercial x-ray image-guidance system for preclinical in vivo studies. To our knowledge this is the first description of an experimental image-guided proton beam for preclinical radiobiology research. It will enable in vivo investigations of radiobiological effects in proton beams.