Environmental Gamma Radiation Monitoring by Means of TLD and Ionisation Chamber

M. Benkrid, D. Mebhah, S. Djeffal, A. Allalou

Radiation Protection Dosimetry · 1992 · 13 citations · 0 references

Concepts

Abstract

In addition to the national radon monitoring programme being carried out by the external dosimetry laboratory of the Health Physics and Safety Centre, a large-scale survey for the monitoring of gamma outdoor radiation is planned. In this context CaSO4:Dy, CaF2:Dy and LiF:Mg,P,Cu have been systematically investigated. CaSO4:Dy is shown to be the best compromise for use in the field experiment. In order to check the passive method based on TL detectors, an active method using a high pressure ionisation chamber (HPIC) type RS-111 has been used. The experimental work was conducted for 1 year in the region of Algiers. The average dose rate in air over the whole monitoring period was found to be 63.50±4.60 nGy.h-1 and 67.70±6.70nGy.h-1, respectively, for the TLDs and the HPIC. A comparison of these two methods showed that these results agree within 7%. The passive method based on the TL detectors can be considered satisfactory and very convenient for a large-scale survey.