Computer Methods in Biomechanics & Biomedical Engineering · 2016 · 35 citations · 28 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBone GraftingImplant SurgeryReconstructive SurgeryBone Defect ReconstructionBone ReconstructionComputer-aided Implant DesignImplant DesignThree-dimensional ReconstructionVirtual Treatment Planning3D Anatomical ReconstructionBone Defect MorphometryCustom Implant DesignCustom Implant ReconstructionCustom Made ImplantsCustom Reconstructive ImplantsThree-dimensional Anatomical ReconstructionThree-dimensional Surgical Simulation
Custom implants are used to treat patients with large acetabular bone defects. To quantify the bone defect and to initialize the implant design, a virtual anatomical reconstruction of the bone needs to be performed. Our SSM-based reconstruction approach was used to overcome the limitations of the mirrored contralateral method and improves upon other SSM reconstruction techniques. The reconstruction errors for the acetabular direction, the hip joint center and the acetabular radius were, respectively: [Formula: see text], 2.6 mm and 0.7 mm. We believe that our method can be an essential tool in the planning and the design of custom implants.
28
Svante Wold, Kim H. Esbensen, Paul Geladi · Chemometrics and Intelligent Laboratory Systems · 1987 · 11.5K citations
Marching cubes: A high resolution 3D surface construction algorithm
William E. Lorensen, H. E. Cline · 1987 · 10.2K citations · Full text
Computing, Radiology, Visualization +17
Gray's anatomy: the anatomical basis of clinical practice
Choice Reviews Online · 2005 · 4.2K citations
Medicine, Anatomy, Neurobiology +17
Generalized Procrustes Analysis
J. C. Gower · Psychometrika · 1975 · 3.2K citations