Biomedical Optics Express · 2016 · 49 citations · 13 references
MedicineAnatomyOtolaryngologyBiomedical ImagingOtologyPediatric OtolaryngologyEar AnatomyEar ReconstructionEar SurgeryOtitis MediaPediatric OtologyTympanic Membrane AnatomyAccuracyCurvatureFeasibility Study DesignField CurvatureIn Vivo ImagingLateral ResolutionThree-dimensional Shape MeasurementTympanic Membrane Shape
We present a light field digital otoscope designed to measure three-dimensional shape of the tympanic membrane. This paper describes the optical and anatomical considerations we used to develop the prototype, along with the simulation and experimental measurements of vignetting, field curvature, and lateral resolution. Using an experimental evaluation procedure, we have determined depth accuracy and depth precision of our system to be 0.05-0.07 mm and 0.21-0.44 mm, respectively. To demonstrate the application of our light field otoscope, we present the first three-dimensional reconstructions of tympanic membranes in normal and otitis media conditions, acquired from children who participated in a feasibility study at the Children's Hospital of Pittsburgh of the University of Pittsburgh Medical Center.
13
Least-squares fitting of circles and ellipses
Walter Gander, Gene H. Golub, Rolf Strebel · BIT Numerical Mathematics · 1994 · 850 citations
Marc Levoy, Ren Ng, Andrew Adams et al. · ACM Transactions on Graphics · 2006 · 737 citations
Gaël Guennebaud · 2007 · 407 citations
Gaël Guennebaud · ACM Transactions on Graphics · 2007 · 331 citations