Optimization of on‐line hydrogen stable isotope ratio measurements of halogen‐ and sulfur‐bearing organic compounds using elemental analyzer–chromium/high‐temperature conversion isotope ratio mass spectrometry (EA‐Cr/HTC‐IRMS)

45

Citations

23

References

2016

Year

Abstract

The optimized EA-Cr/HTC reactor design can be implemented in existing analytical equipment using commercially available material and is universally applicable for both heteroelement-bearing and heteroelement-free organic-compound classes. The sensitivity and simplicity of the on-line EA-Cr/HTC-IRMS technique provide a much needed tool for routine hydrogen-isotope source tracing of organic contaminants in the environment. Copyright © 2016 John Wiley & Sons, Ltd.