Numerical Simulation of 4K GM Cryocooler

Mingzhi Xu, Takaaki Morie

2012 · 13 citations · 1 references

Concepts

Abstract

A numerical simulation program has been developed to predict the theoretical cooling capacity and losses in regenerators, heat exchangers, etc. The clearance between the cylinder and the displacer is modeled as a pulse tube in a pulse tube cryocooler. Usually, a slip seal is installed at the warm end of the first stage. Therefore, the clearance is calculated as a pulse tube in a basic pulse tube cryocooler. On the other hand, there is no seal on the warm end of the second stage. Therefore, the clearance is calculated as a pulse tube in a basic double inlet pulse tube cryocooler. The spiral groove on the second stage displacer is also calculated as a pulse tube in a basic double inlet pulse tube cryocooler. The cooling effect in the pulse tube can be calculated by assuming that the pulse tube is adiabatic. In other words, there is no heat exchange between gas and the walls of the cylinder and the displacer. Since the volume of both the clearance and the spiral groove is so small that the cooling effect in the pulse tube is small. The simulation results of the pressure and volume variation, and the P-V power in the second stage expansion volume are compared with the measurement results obtained from an SHI 4K GM cryocooler. The simulation results are consistent with the measurement results.