Microwave breast imaging (MBI) uses low power and longer wavelength signals (compared to X-ray mammography) to obtain information about breast tissues, and promises a safer and more accurate modality for regular breast scanning. One of the major impediments of this technology has been a large signal re∞ection from the breast skin. In this paper, we present a method to reducing the efiect of signal re∞ection from the breast skin by placing the antenna in-contact with the breast skin. In the reported method, the skin is considered a layer of the antenna substrate, and the efiect of having the antenna in contact with the skin is included in the antenna design itself. Thus, the design allows placing the antenna on the breast skin. This reduces the signal scattering from the skin and more transmitted signal is irradiated on the tumor, thus, increasing the tumor detection sensitivity. Design and simulation in Ansoft High Frequency Simulation Software (HFSS) is presented. The simulation results show that the current density in the breast skin is much less while the current density in the tumor is much higher when the antenna is placed directly on the breast skin, compared to when the antenna is placed away from the skin. While, in this paper, the antenna is discussed in the context of microwave breast imaging, the concept of placing an antenna directly on human skin has a wide range of biomedical applications.

References

8