Concepedia

Publication | Open Access

F <sub> IO <sub>2</sub> </sub> in an Adult Model Simulating High-Flow Nasal Cannula Therapy

54

Citations

14

References

2016

Year

Abstract

During HFNC with simulated spontaneous breathing, when HFNC gas flow was 60 L/min, measured F<sub>IO<sub>2</sub></sub> was similar to set F<sub>IO<sub>2</sub></sub> at 0.3 and 0.5, whereas at 0.7, as V<sub>T</sub> increased, measured F<sub>IO<sub>2</sub></sub> decreased slightly. However, at 20 or 40 L/min, changes in V<sub>T</sub> related with deviation from set F<sub>IO<sub>2</sub></sub> .

References

YearCitations

Page 1