The 36-T Series-Connected Hybrid Magnet System Design and Integration

Iain R. Dixon, S. Bole, K.R. Cantrell, S. T. Hannahs, John Kynoch, William Marshall, A. Powell, J. Tóth, M.D. Bird

IEEE Transactions on Applied Superconductivity · 2016 · 12 citations · 7 references

Concepts

Abstract

The system integration of a 36 T high field resistive-superconducting hybrid magnet system which will be used primarily for NMR studies is being finalized at the National High Magnetic Field Laboratory in Tallahassee, Florida. The hybrid magnet consists of a 23 T resistive insert coil set nested inside of a 13 T superconducting coil wound with high J <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">C</sub> Nb <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> Sn/Cu cable-in-conduit conductor. The resistive and superconducting coils are connected electrically in series and operate at 20 kA. The combined system will enable 1-ppm level uniformity in a 1 cm DSV for condensed-matter NMR at a record level of 1.53 GHz. The resistive insert has four concentric Florida-Bitter coils and operates at 14 MW. The superconducting outsert operates with forced flow supercritical helium at 4.5 K and 3.5 bar. The superconducting protection system consists of two fully independent circuits, one employing voltage averaging comparisons between winding layers and the other a voltage comparison with a co-wound coil. Initial magnet testing of the resistive coils has been completed. The superconducting cold mass has been cooled to 4.5 K and the system is being prepared for combined hybrid magnet testing.

References

7