NeuroImage · 2016 · 95 citations · 50 references
MedicineNeuroscienceRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance SpectroscopyMedical ImagingMedical Imaging Instrumentation3D Magnetic Resonance ImagingBrain RegionsFree Induction DecayHigh-resolution Magnetic Resonance ImagingNeuroscience ImagingParallel ImagingParallel Imaging SequencesPulse Sequence Design
in a comparatively short measurement time. A clinically feasible scan time of 10-20min is reached via a short TR of 200 ms due to an optimised free induction decay-based acquisition with shortened water suppression as well as parallel imaging (PI) using Controlled Aliasing In Parallel Imaging Results IN Higher Acceleration (CAIPIRINHA). This approach is not limited to a rectangular region of interest in the centre of the brain, but also covers cortical brain regions. Transversal pulse-cascaded Hadamard encoding was able to further extend the coverage to 3D-UHR-MRSI of four slices (100×100×4 matrix size), with a measurement time of 17min. Lipid contamination was removed during post-processing using L2-regularisation. Simulations, phantom and volunteer measurements were performed. The obtained single-slice and 3D-metabolite maps show the brain in unprecedented detail (e.g., hemispheres, ventricles, gyri, and the contrast between grey and white matter). This facilitates the use of UHR-MRSI for clinical applications, such as measurements of the small structures and metabolic pathologic deviations found in small Multiple Sclerosis lesions.
50
Philip M. Robson, Aaron K. Grant, Ananth J. Madhuranthakam et al. · Magnetic Resonance in Medicine · 2008 · 434 citations · Full text