Performance evaluation of a prototype Positron Emission Tomography scanner using Digital Photon Counters (DPC)

46

Citations

3

References

2012

Year

Abstract

We show performance results of a prototype Positron Emission Tomography scanner based on digital SiPMs, or Digital Photon Counters (DPC), developed by the Philips Digital Photon Counting unit. The scalability of the DPC technology is demonstrated by an excellent system coincidence timing resolution of 266 ps FWHM and an energy resolution of 10.7 % FWHM. Even while using 4 mm × 4 mm × 22 mm LYSO crystals, the spatial resolution is close to 2.4 mm. Although not optimized yet, the image homogeneity is 5.8 %. We show that the system performance is maintained even at highest count rates encountered in PET scans.