MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingDetector TechnologyImagingMagnetic Resonance ImagingMedical ImagingRadiation Imaging DetectorsDetectorsImaging DetectorsMri CompatibilityPinhole DetectionSingle-photon Emission Computed Tomography Instrumentation
This paper presents the design study of an MRI compatible, ultra-high resolution SPECT system. This system consists of multiple ultra-high resolution energy-resolved photon counting (ERPC) CdTe detectors that are currently development. This detector is based on a novel readout ASIC that is roughly 1.1x2.2cm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> in size and has a 32x64 array of 350x350um <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> pixels. A prototype detector was tested inside a 4.7 Tesla MRI scanner for simple pinhole imaging. The results obtained have demonstrated the feasibility of simultaneous acquisition of SPECT and MRI data. Extensive Monte Carlo simulation was performed to evaluate a number of system configurations.
3
Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI.
Bernd J. Pichler, Martin S. Judenhofer, Ciprian Catana et al. · PubMed · 2006 · 397 citations