A prototype of a novel transformable single-ring OpenPET

Taiga Yamaya, Eiji Yoshida, Hideaki Tashima, Yasunori Nakajima, Fumihiko Nishikido, Yoshiyuki Hirano, Naoko Inadama, Hiroshi Ito, Tetsuya Shinaji, Hideaki Haneishi,

2013 · 11 citations · 8 references

Concepts

Abstract

We are developing the world's first, open-type 3D PET scanner “OpenPET” for PET-image guided particle therapy such as in situ dose verification and direct tumor tracking. Following our first idea of a dual-ring Open PET (DROP), we proposed our second-generation geometry, single-ring OpenPET (SROP), which is more efficient than DROP in terms of manufacturing cost and sensitivity. In this paper, we have developed a SROP prototype based on a novel detector arrangement, in which block detectors originally forming a conventional PET scanner were axially shifted little by little. Sixteen detector units each of which consists of two depth-of-interaction detectors are arranged to form a perfect circle, 25cm in diameter. Detector units have an axial shifting mechanism so that they can be transformed into the SROP; adding this mechanism to the units allows us to use the scanner as a conventional (i.e., non-open) PET when in-beam PET measurements are not required. After confirming its basic imaging performance using a phantom filled with <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">18</sup> F solution, we carried out in-beam imaging tests in the Heavy Ion Medical Accelerator in Chiba (HIMAC). In addition to the usual carbon ( <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">12</sup> C) beam, we applied RI beams of <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">11</sup> C and <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">10</sup> C. Stopping positions of primary particles were directly imaged with the RI beam irradiation, while the stopping position distribution of secondary particles was imaged with the <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">12</sup> C beam irradiation. Phantom study results with pencil beam irradiation of about 2.5Gy showed that beam stopping positions can be measured with the precision better than 2mm with the <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">11</sup> C beam irradiation followed by 20 min PET measurement. With the <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">10</sup> C beam, PET measurement time could be reduced to 1/10 while still maintaining the precision. For both <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">11</sup> C and <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">10</sup> C, there is room for further reduction of PET measurement time.

References

8