2002 · 63 citations · 4 references
EngineeringBiomedical EngineeringBioinstrumentationElectrophysiologyNeural EngineeringEvoked PotentialsImplant TechnologyNeural Signal AcquisitionNeurostimulationClosed-loop SystemsElectrode-nerve InterfaceElectrophysiological RecordingsMechanical StimulationNeural Signal AmplificationRecordingsSignals
Recording cuffs used to register the activity of sensory afferents in peripheral nerves hold promise to provide feedback signals for closed-loop FES systems. This paper compares two modes in which the electrodes of a tripolar nerve cuff may be connected to amplifier components. The authors recorded evoked sensory activity from the sciatic nerve in an anesthetized rabbit while mechanical stimulation was applied to the central toe pad. A commercial transformer coupled amplifier and a custom amplifier that uses AMPO1 integrated circuit components were compared when each amplifier was configured in a 'quasi-tripolar' verses true tripolar configuration. Results showed that the true tripolar configuration had superior signal to noise ratio in both amplifiers. Furthermore, an excellent performance was demonstrated for the AMPO1 based amplifier without transformer coupling, suggesting that amplifiers designed for implantation (which precludes using bulky transformers) need not compromise the quality of the neural recordings.
4
Stable long-term recordings from cat peripheral nerves
R. B. Stein, T. Richard Nichols, Jack H. Jhamandas et al. · Brain Research · 1977 · 161 citations
Restoration of Lateral Hand Grasp Using Natural Sensors
Morten Kristian Haugland, A. Lickel, Ron Riso et al. · Artificial Organs · 1997 · 52 citations · Full text