Concepedia

Publication | Closed Access

Biomechanical Comparison of Expansive Pedicle Screw and Polymethylmethacrylate-augmented Pedicle Screw in Osteoporotic Synthetic Bone in Primary Implantation

31

Citations

35

References

2013

Year

Abstract

EPS can significantly increase the strength of screw fixation compared with CPS in osteoporotic synthetic bone. Although EPS shows a weaker fixation strength compared with PMMA-PS in the osteoporotic synthetic bone, it may still provide an alternative option to prevent screw loosening in the clinical treatment of osteoporosis.