Concepedia

Publication | Closed Access

Invasive validation of a novel brachial cuff-based oscillometric device (SphygmoCor XCEL) for measuring central blood pressure

72

Citations

21

References

2016

Year

Abstract

The impaired accuracy of central BP measurement was mainly due to calibration-derived, but not device-dependent, bias. Strong correlation between oscillometric and invasive measurements indicates that SphygmoCor XCEL warrants future investigations to determine the clinical validity of this device.