A neuro-stimulus chip with telemetry unit for cochlear implant

Man Dong, Chun Zhang, Zhihua Wang, Dongmei Li

2005 · 17 citations · 4 references

Concepts

Abstract

This paper presents a 16-channel cochlear implant which supports 4 stimulus modes and provides a highest stimulus rate of 15k pulses per second per channel, and dynamic range of 60dB with the 10 bit D/A converter. The device consists of a coil, a rectifier and regulator, and a mixed analog and digital neuro-stimulus ASIC which decodes telemetry command and generates stimulus signals. The chip is fabricated using 1.2/spl mu/m 2M/2P standard CMOS process and the die size is 3.56 mm /spl times/ 3.67mm. The circuit is working properly at a telemetry data rate of 500 Kbps in carrier of 10 MHz while the peak power consumption is 35 mW.

References

4