MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyAbdominal ImagingImage QualityImage Quality AssessmentImage Quality AssuranceMedical ImagingRadiographic ImagingClinical Image AssessmentClinical Image QualityImaging System CalibrationRadiographic Image QualityWater Phantom Measurements
and SNR did, with the latter being the preferred calibration metric given that it is easier to measure in practice. This result was consistent with the previous investigation for chest imaging with AEC devices. Medical physicists may therefore use a simple and easily accessible uniform water equivalent phantom to measure the SNR image quality metric described here when calibrating AEC devices for abdomen, pelvis and spine imaging with Agfa CR systems, in the confidence that clinical image quality will be sufficient for the required clinical task. However, to ensure appropriate levels of detector air kerma the advice of expert image evaluators must be sought.