Theoretical Study and Calculation of the Response of a Fast Neutron Dosemeter Based on Track Detection

J.L. Decossas, Jean-Claude Vareille, J.P. Moliton, J.L. Teyssier

Radiation Protection Dosimetry · 1983 · 14 citations · 0 references

Concepts

Abstract

At the present time, a fast neutron dosemeter is generally composed of a radiator in which n-p elastic scattering occurs and a detector which registers protons. The theoretical study, and the calculation (FORTRAN programme) of the response of such a dosemeter which we present here involves two steps: (1) The proton flux emerging from a thick radiator on which monoenergetic neutrons are normally incident is studied. We characterise it by its energy spectrum depending on the neutron energy and on the radiator thickness. (2) Proton detection being achieved with a solid state nuclear track detector whose performance is known, we calculate the number of registered tracks. The dosemeter sensitivity (tracks cm-2.Sv-1) is deduced. Then, our calculations show that it is possible to optimise the radiator thickness to obtain the smallest variation in sensitivity with neutron energy. Our theoretical results are in good agreement with the experimental ones found in the literature.