Practical Laboratory Medicine · 2016 · 22 citations · 6 references
MedicineHematologyAnalytical ChemistryHemostasisInstrumentationBioanalytical Method ValidationClinical HematologyCoagulation DisordersHemostatic InstrumentsMedical InstrumentationCoagulase TestCoagulationCorrelation AnalysisError Rate AnalysisImmunological AssaysLaboratory Diagnosis Of CoagulopathyOperational Performance AssessmentThroughput
OBJECTIVES: System™ (Siemens) for analytical performance. Additionally, the operational performance of the Sysmex CS-5100 System was compared with the blood coagulation analyzer ACL TOP 700 (Instrumentation Laboratory, Werfen Group, Kirchheim bei Munchen, Germany). MATERIALS AND METHODS: or the ACL TOP in the measurement of patients' samples. RESULTS: The inter-assay precision using the CS systems was impressive (inter-assay CV generally <3.5%) and the correlation between the two Sysmex analyzers was excellent. In the throughput study, the CS-5100 completed the measurement of 100 samples (210 results) in less than 49 min. CONCLUSIONS: Our results demonstrated that the CS-5100 is a robust high-throughput analyzer, well-suited for coagulation laboratories.
6
Goce Dimeski · PubMed · 2008 · 47 citations
Evaluation of the automated coagulation analyzer Sysmex® CA-7000
Alexandra Dorn‐Beineke, Carl-Erik Dempfle, Thomas Bertsch et al. · Thrombosis Research · 2005 · 27 citations
Evaluation of the Automated Coagulation Analyzer SYSMEX CA 6000
Peter Quehenberger, Stylianos Kapiotis, Sylvia Handler et al. · Thrombosis Research · 1999 · 19 citations
Medicine, Hematology, Hemostasis +8