EngineeringChemical EngineeringMaterials ScienceBiosensorsElectrochemistrySensorsElectrochemical BiosensorsElectrochemical DetectionElectrochemical SensorsMolecular SensorsNanomaterial-based BiosensorsNanosensorsBiochemical SensingElectrochemical Molecular SensorsElectrochemical NanosensorsElectrochemical SensingMolecularly Imprinted Polymer SensorsNanomaterial-modified ElectrodesPolymer-based Biosensors
This work used a highly selective electrochemical sensor based on polythionine/AuNPs composites (PTh/AuNPs) modified glassy carbon electrode (GCE) to detect epinephrine (EP). An elevated response of the oxidative behavior of EP was obtained on the PTh/AuNPs modified GCE in phosphate buffer saline (pH 6.0), which allowed sensitive sensing of EP using the differential pulse voltammetry (DPV). The electrochemical sensor can selectively measure EP in a wide linear range from 1 to 40 mg/L with a detection limit of 0.3 mg/L (S/N = 3) and a high reproducibility in PBS. The developed electrochemical sensor can be favorably applied to detect EP in a diluted serum. Furthermore, the good linear regression between the EP concentrations and the currents in various diluted serum implied the potential of the proposed method as a promising candidate approach for complex clinical conditions.