Experimental Setup to Compare Measurements and Numerical Simulations in Magnetic Resonance Imaging RF Dosimetry

Umberto Zanovello, M. Borsero, Domenico Giordano, Luca Zilberti, Francesca Maggiorelli, Gianluigi Tiberi

IEEE Transactions on Instrumentation and Measurement · 2017 · 11 citations · 5 references

DOIFull text

Open access

Concepts

Abstract

Many of the parameters associated with the magnetic resonance imaging (MRI) coils are estimated by means of numerical simulations. Taking into account the unavoidable numerical approximations and imperfections of the models, an experimental validation of the theoretical results becomes essential. This paper describes two measuring setups which allow the comparison between measurements of electromagnetic fields and the same quantities computed numerically. The experimental activity highlighted some critical aspects of the numerical results that could bring to a wrong estimation of the parameters associated with the MRI coils. Results show the importance and feasibility of a dosimetry experimental setup suitable for MRI coils characterization.

References

5