Radiocarbon · 1980 · 11 citations · 2 references
EngineeringNuclear EngineeringInstrumentationNuclear InstrumentationRadiation ShieldingBackground Radiation MeasurementBackground ReductionRadiation DetectorsScintillation DetectorsWater ShieldingBackground Radiation TestingLaboratory Background MeasurementLarge-volume DetectorsLiquid Scintillation CountingLiquid ScintillatorsScintillationScintillation DetectionScintillation Yield
A commercial liquid scintillation counter (LSC) was tested for its ability of low level counting (LLC) in two underground laboratories (70m of water equivalent cover). Reduction of total background and background components compared to the normal counting laboratory (5m w.e. cover) up to a factor of 9 depending on the energy range and the shielding of counting cell and laboratory walls were obtained. First experiments with a large volume cell (50cc) led to a modern 14 C net effect of 425cpm and a corresponding background of 1.6cpm.
2
H. Oeschger, M. Wahlen · Annual Review of Nuclear Science · 1975 · 37 citations