Standardization of Ni-63 by 4 pi beta liquid scintillation spectrometry with H-3-standard efficiency tracing

Brian E. Zimmerman, R. Collé

Journal of Research of the National Institute of Standards and Technology · 1997 · 41 citations · 2 references

DOIFull text

Open access

Concepts

Abstract

= 2 and thus a 2 standard deviation estimate) uncertainty based on the evaluation of 28 different uncertainty components. These components were evaluated on the basis of an exhaustive number (976) of LS counting measurements investigating over 15 variables. Through the study of these variables it was found that LS cocktail water mass fraction and ion concentration play important roles in cocktail stability and consistency of counting results. The results of all of these experiments are discussed.

References

2