Progress at the Isotrace Radiocarbon Facility

R.P. Beukens, Debbie Gurfinkel, Henry W Lee

Radiocarbon · 1986 · 42 citations · 1 references

DOIFull text

Open access

Concepts

Abstract

Since August 1984 over 100 samples have been analyzed. 12 C, 13 C, and 14 C isotopes were measured after acceleration and all dates were corrected for natural and preparation-induced fractionation. A precision of better than 1.0% was obtained for samples 10,000 yr BP or less and these results were shown to be reproducible at precision levels as low as 0.3%. An accuracy of better than 1.0% was demonstrated for the same age range by analyzing samples which were previously dated with the conventional technique. The machine background, measured on blank sample holders in a clean ion source, yielded a count rate equivalent of 0.015 ± 0.007% modern.

References

1