Measurement of thermal conductivity of fluids using 3-ω method in a suspended micro wire

R Karthik, N. Harish Nagarajan, B. Raja, P. Damodharan

Journal of Engineering Thermophysics · 2012 · 10 citations · 11 references

Concepts

Abstract

A thermally controlled, compact device employing the 3- ω technique, used to measure the thermal conductivity of fluids, is designed, developed, and presented in this paper. The 3- ω method, which analyzes temperature oscillations data in the frequency domain, requires a microscopic sample and extremely low heating power. The functionality is derived from the approximate solutions of temperature oscillations of a line heater based on the infinite line-heater model over an empirically and analytically chosen range of frequencies. The method is devoid of errors related to transient measurements, fluid thermal stratification and mobility errors, which pose difficulties in other methods. A platinum (99.99% pure) wire of 50 µm diameter and a length of 30 mm, suspended in a sample volume of 25 µl of the test fluid, serves simultaneously as the heater and thermometer. Structure-wise, the device is designed to support measurements over a range of temperatures and fluid pressures providing modularity and flexibility to the instrument. The device is successfully employed to measure the thermal conductivity of de-ionized water for temperatures between 15 and 35°C with an accuracy of ±1.2% inmeasurement.

References

11