Thermal Collateral Damage in Porcine Corneas After Photoablation With Free Electron Laser

Thomas Bende, Rudolf Walker, B. Jean

Journal of Refractive Surgery · 1995 · 21 citations · 18 references

Concepts

Abstract

BACKGROUND: The study describes a quantitative and systematic investigation of the collateral thermal damage during infrared photoablation as a function of wavelength between 2.7 and 6.7 microns. METHODS: Using the tunable Free Electron Laser at Vanderbilt University, Nashville, Tenn, 60 freshly removed porcine cadaver eyes were irradiated at wavelengths between 2.7 and 6.7 microns, at a fluence of 1.3 J/cm2. For wavelengths where no photoablation occurred, fluence was increased to 3.5 J/cm2; pulse length (macropulse) was 4 microseconds, consisting of a train of micropulses (pulse duration 2 ps at a 2.9 GHz repetition rate). The corneal buttons were removed and stained with hematoxylin and eosin (H&E) and analyzed by histologic micrometry. RESULTS: Two thermal damage zones in the remaining tissue were observed: zone 1 showed superficial carbonization and measured between 2 and 4 microns; beyond, the eosinophilic zone 2 measured between 10 and 100 microns. The extent of zone 2 was inversely related to the absorption spectra of the cornea; it was minimal at the 3- and 6-micrometer water absorption bands and maximal at minimal target absorption. CONCLUSION: The results correlated well with a model of the ablation process. The study provides a systematic and predictive element for the determination of collateral thermal adverse effects; it does not yet include pulse length variation as a determining factor.

References

18