Journal of Healthcare Engineering · 2016 · 110 citations · 26 references
EngineeringHealth InformaticsBioinstrumentationBiomedical Signal ProcessingSensorsCardiac Signal ProcessingCardiovascular MonitoringMedical InstrumentationVital SignsVital Signs RecordingWearable Health TechnologiesCardiac Monitoring SystemsHeart RateHeart Rate EvaluationHeart Rate MonitoringPhysiological MonitoringPhysiological Signal ProcessingStress DetectionStress Monitoring
The automated detection of stress is a central problem for ambient assisted living solutions. The paper presents the concepts and results of two studies targeted at stress detection with a low cost heart rate sensor, a chest belt. In the device validation study ( n = 5), we compared heart rate data and other features from the belt to those measured by a gold standard device to assess the reliability of the sensor. With simple synchronization and data cleaning algorithm, we were able to select highly (>97%) correlated, low average error (2.2%) data segments of considerable length from the chest data for further processing. The protocol for the clinical study ( n = 46) included a relax phase followed by a phase with provoked mental stress, 10 minutes each. We developed a simple method for the detection of the stress using only three time-domain features of the heart rate signal. The method produced accuracy of 74.6%, sensitivity of 75.0%, and specificity of 74.2%, which is impressive compared to the performance of two state-of-the-art methods run on the same data. Since the proposed method uses only time-domain features, it can be efficiently implemented on mobile devices.
26
Paolo Melillo, Marcello Bracale, Leandro Pecchia · BioMedical Engineering OnLine · 2011 · 326 citations · Full text