Japanese Journal of Radiological Technology · 2004 · 21 citations · 3 references
Veterinary MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImage QualityImage Quality AssessmentMedical ImagingPatient Radiation ProtectionDose Reduction TechniquesImage Acquisition QualityLocal ExposureLow-dose Computed TomographyPatient Radiation Dose OptimizationRadiation Dose ManagementRadiation Dose OptimizationRadiation Dose Reduction
CT perfusion study requires repeatedly scanning the same part of the patient's head, resulting in an increase of local radiation exposure. The purpose of this study was to assess the feasibility of the ultra-low-dose technique with a quantum de-noising filter. The newly developed quantum de-noising filter selectively reduced noise by two-thirds, while maintaining spatial resolution. The low-dose protocol using the quantum de-noising filter and slow rotation speed accomplished a 68% reduction in the local radiation dose compared with the previous standard protocol. The quantum de-noising filter is considered to be useful to reduce radiation exposure and to improve image quality in CT perfusion study.
3
Using 80 kVp versus 120 kVp in perfusion CT measurement of regional cerebral blood flow.
Max Wintermark, Philippe Maeder, Francis R. Verdun et al. · PubMed · 2001 · 252 citations · Full text
Kohsuke Kudo, Satoshi Terae, Chietsugu Katoh et al. · PubMed · 2003 · 213 citations · Full text