Design and Clinical Evaluation of a Handheld Wavefront Autorefractor
34
Citations
22
References
2015
Year
The study introduces a low‑cost, handheld QuickSee autorefractor designed to measure refractive errors without moving parts or expensive components. The authors recruited 41 adults to perform noncycloplegic refraction with the QuickSee prototype, a Grand Seiko autorefractor, and subjective refraction, and compared the devices using Bland‑Altman analysis. Both autorefractors showed strong correlation with subjective refraction (r = 0.97) and agreement within ±0.84 D; the QuickSee’s spherical and cylindrical accuracy matched the Grand Seiko’s within 0.25 D in 74 % and 87 % of eyes, indicating equivalent performance.
To introduce a novel autorefractor design that is intended to be manufacturable at low cost and evaluate its performance in measuring refractive errors.We developed a handheld, open-view autorefractor (the "QuickSee" [QS]) that uses a simplified approach to wavefront sensing that forgoes moving parts and expensive components. Adult subjects (n = 41) were recruited to undergo noncycloplegic refraction with three methods: (1) a QS prototype, (2) a Grand Seiko WR-5100K (GS) autorefractor, and (3) subjective refraction (SR). Agreements between the QS and GS were evaluated using a Bland-Altman analysis. The accuracy of both autorefractors was evaluated using SR as the clinical gold standard.The spherical equivalent powers measured from both autorefractors correlate well with SR, with identical correlation coefficients of r = 0.97. Both autorefractors also agree well with each other, with a spherical equivalent power 95% confidence interval of ±0.84 diopters (D). The difference between the accuracy of each objective device is not statistically significant for any component of the power vector (p = 0.55, 0.41, and 0.18, for M, J0, and J45, respectively). The spherical and cylindrical powers measured by the GS agree within 0.25 D of the SR in 49 and 82% of the eyes, respectively, whereas the spherical and cylindrical powers measured by the QS agree within 0.25 D of the SR in 74 and 87% of the eyes, respectively.The prototype autorefractor exhibits equivalent performance to the GS autorefractor in matching power vectors measured by SR.
| Year | Citations | |
|---|---|---|
Page 1
Page 1