Calibration of Contactless Pulse Oximetry

90

Citations

15

References

2016

Year

Abstract

For healthy adults, the results present strong evidence that camera-based contactless pulse oximetry is fundamentally feasible because long-term (eg, 10 minutes) error stemming from variation among individuals expressed as A*rms is significantly lower (<1.65%) than that required by the International Organization for Standardization standard (<4%) with the notion that short-term errors should be added. A first illustration of such errors has been provided with A**rms = 2.54% for 40 individuals, including 6 with dark skin. Low signal strength and subject motion present critical challenges that will have to be addressed to make camera-based pulse oximetry practically feasible.