Journal of AOAC INTERNATIONAL · 1982 · 15 citations · 0 references
PharmacyPharmaceutical AnalysisChromatographic AnalysisDrug MetabolismMass Spectrometric AnalysisAnalytical ChromatographyChromatographyDrug DispositionLiquid ChromatographyMetabolite ExtractionMetabolite IdentificationHigh-performance Liquid Chromatography AnalysisInternal Standard AnalysisInternal StandardizationMetabolite Extraction ProtocolsMetabolite Separation TechniquesStandardization
A high pressure liquid chromatographic technique with internal standardization has been developed for determining zearalenone and metabolites in rat urine and liver. Following extraction with methylene chloride and solvent partition, samples are cleaned up by applying the extract to a Sephadex LH-20 column and eluting with a mixture of benzene-methanol (85 + 15). Compounds were resolved on 2 Part-isil-10 columns (25 cm x 4.6 mm id) in series with a mobile phase of isooctane-chloroform-methanol (35 + 25 + 3), and detected at 280 nm. The internal standard was 6'alpha-acetoxyzearalane. Limits of detection were about 2.0 ng for zearalenone and 5.0 ng for zearalenols (6'-hydroxyzearalane). Zearalenone and zearalenols were excreted mainly in free form with relatively little glucuronide conjugation. Metabolism of zearalenone to free zearalenol was minor compared with formation of bound forms.